| ... | ... | @@ -359,7 +359,7 @@ mydata.header[3].UB # return dictionary of UB metadata |
|
|
|
## Troubleshooting
|
|
|
|
|
|
|
|
1. There is no X-ray beam in Kappa.
|
|
|
|
- Make sure there is beam in the storage ring.
|
|
|
|
- Make sure there is current in the storage ring.
|
|
|
|
- Check if main shutter and branch shutter are open. If any shutter cannot be opened, Check beamline EPS and contact beamline staff.
|
|
|
|
- Check if ID energy and mono energy setpoints are consistent.
|
|
|
|
- Check beam on the exit slit camera (cam6) on the m3r_align panel. If there is no beam and piezo voltage is very high (-12V or 40V), try set it back to zero.
|
| ... | ... | |