## Troubleshooting
|
|
## Troubleshooting
|
|
|
|
|
|
|
|
1. There is no beam.
|
|
1. There is no X-ray beam in Kappa.
|
|
|
- Make sure there is beam in the storage ring.
|
|
- Make sure there is beam in the storage ring.
|
|
|
- Check if main shutter and branch shutter are open. If any shutter cannot be openned, Check beamline EPS and call staff.
|
|
- Check if main shutter and branch shutter are open. If any shutter cannot be opened, Check beamline EPS and contact beamline staff.
|
|
|
- Check if ID and mono energy setpoints are consistent.
|
|
- Check if ID energy and mono energy setpoints are consistent.
|
|
|
- Check beam on the exit slit camera on the m3r_align panel.
|
|
- Check beam on the exit slit camera (cam6) on the m3r_align panel. If there is no beam and piezo voltage is very high (-12V or 40V), try set it back to zero.
|
|
|
- If there is no beam and piezo voltage is very high (-12V or 40V), try set it back to zero.
|
|
|
|
|
|
|
|
|
|
2. `m3r_align` doesn't work.
|
|
2. `m3r_align` doesn't work.
|
|
|
- This happens when the code had hard time to determine where the peak is. Try tweaking the piezo voltage back to zero and adjust m3r pitch.
|
|
- This happens when the code had hard time to determine where the peak is. Try tweaking the piezo voltage back to zero and adjust m3r pitch.
|
|
|
- When the photon energy is too high (> 2.3 keV), or the mono and ID energies are very different, photon counts can be too low to determine the peak position. Try increasing gain and exposure time (`t(s)` on the screen) of the camera.
|
|
- When the photon energy is too high (> 2.3 keV), or the mono and ID energies are very different, photon counts can be too low to determine the peak position. Try increasing gain and exposure time (`t(s)` on the screen) of the camera.
|
|
|
|
|
|
|
|
3. ID gets stuck.
|
|
3. ID gets stuck.
|
|
|
- Change mode on the ID screen (29ID main screen -> ID -> IEX Machine Physics), wait until it's done, and change the mode back. This will reset the ID.
|
|
- Change mode on the ID screen (29ID main screen -> ID -> IEX Machine Physics), wait until it's done and change the mode back. This will reset the ID.
|
|
|
|
|
|
|
|
4. Motor gets stuck.
|
|
|
|
- Turn on the light in Kappa chamber and check if there is any collision or cable winding.
|
|
|
|
- th and chi motors tend to get sticky after scanning around a peak for many times. Try tweaking the motor back and forth manually.
|
|
|
|
- Contact beamline staff if moving back and forth doesn't help.
|
|
|
|
|
|
|
|
5. Motor positions are incorrect.
|
|
|
|
- Turn on the light in Kappa chamber and check motor positions.
|
|
|
|
- If positions do not reflect physical positions. This is usually due to ioc crash/reboot. Contact beamline staff.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ... | | ... | |