___
|
|
|
___
|
|
|
|
|
|
# Content
|
|
|
# content
|
|
|
|
|
|
- [coming out of shutdown](#coming-out-of-shutdown)
|
|
|
- [energy calibration](#energy-calibration)
|
|
|
- [align M3R pitch to center of rotation](#align-m3r-pitch-to-center-of-rotation)
|
|
|
- [determine kth_offset](#determine-kth_offset)
|
|
|
- [align M4R pitch to center of rotation](#align-m4r-pitch-to-center-of-rotation)
|
|
|
- [determined kap offset](#determined-kap-offset)
|
|
|
- [determined kphi and phi 0](#determined-kphi-and-phi-0)
|
|
|
- [determined kap offset](#determine-kap-offset)
|
|
|
- [determined kphi and phi 0](#determine-kphi-and-phi-0)
|
|
|
|
|
|
___
|
|
|
___
|
|
|
|
|
|
|
|
|
## Coming out of shutdown
|
|
|
## coming out of shutdown
|
|
|
1) home motors using procedure in [elog](https://logbook.xray.aps.anl.gov/29id/RSXS+Endstation/44)
|
|
|
2) check tth minimum/maximum allowed position in extended panel (PI software): after homing tth PI/dial position will be set to either of those, which will prevent it to move it in that direction
|
|
|
3) sync EPICS values:
|
... | ... | @@ -139,7 +139,7 @@ align_z0_chi0() |
|
|
|
|
|
---------------------------------------
|
|
|
|
|
|
## determined kap offset
|
|
|
## determine kap offset
|
|
|
|
|
|
|
|
|
1) compare kap scans at (th,chi,phi) = (90,0,90) and (270,0,-90):
|
... | ... | @@ -154,7 +154,7 @@ dscankap(-4,4,0.1) |
|
|
|
|
|
---------------------------------------
|
|
|
|
|
|
## determined kphi and phi 0
|
|
|
## determine kphi and phi 0
|
|
|
|
|
|
<ol>
|
|
|
<li>
|
... | ... | |