... | @@ -4,8 +4,18 @@ How to home tth from ![image](uploads/980f2f9c2666873734aa9fac9361a535/image.png |
... | @@ -4,8 +4,18 @@ How to home tth from ![image](uploads/980f2f9c2666873734aa9fac9361a535/image.png |
|
|
|
|
|
- enter IP address: 164.54.118.24
|
|
- enter IP address: 164.54.118.24
|
|
- turn current on: ![image](uploads/32cfab7202eaab25ab23f014faf2b804/image.png) => Motor On + send
|
|
- turn current on: ![image](uploads/32cfab7202eaab25ab23f014faf2b804/image.png) => Motor On + send
|
|
- check soft limits
|
|
- check soft limits: ![image](uploads/33c6c1d2a222252947b903e610567405/image.png)
|
|
|
|
- set limit settings to ignore normally open x2: ![image](uploads/22033937627b72b32825c700066dc509/image.png)
|
|
|
|
![image](uploads/650417152f86e41f8ac51341651efdca/image.png)
|
|
|
|
- move between the 2 limits: they about 10 degrees apart, a little after direct beam on the negative hemisphere of the chamber
|
|
|
|
- E1 = cal = limit reverse
|
|
|
|
- E2 = rm = limit forward
|
|
|
|
- ![image](uploads/1a6b0405009c84853cbd395acc9ad865/image.png)
|
|
|
|
- set limit settings to normally closed x2
|
|
|
|
- once in between the 2 limits, click ncal ![image](uploads/e40cef0f3899e5828930152ac30811f6/image.png)
|
|
|
|
- software should reset tth to 0
|
|
|
|
- reset soft limits
|
|
|
|
- reset limit settings to ignore normally open x2
|
|
|
|
|
|
|
|
|
|
|
|
|
... | | ... | |