| ... | @@ -11,14 +11,14 @@ To home tth from : m |
... | @@ -11,14 +11,14 @@ To home tth from : m |
|
|
- In pop up screen, enter IP address: 164.54.118.24
|
|
- In pop up screen, enter IP address: 164.54.118.24
|
|
|
- Turn current on:  => Motor On + send
|
|
- Turn current on:  => Motor On + send
|
|
|
- Limit switches are close to the position where d4 is in direct beam. Check soft limits allow movement over that range: 
|
|
- Limit switches are close to the position where d4 is in direct beam. Check soft limits allow movement over that range: 
|
|
|
- Set limit settings to ignore normally closed for both directions: 
|
|
- Set limit settings to ignore normally closed for both directions in: 
|
|
|

|
|

|
|
|
- Move between the 2 limits and look for when the limit light flashes: They are about 10 degrees apart, a little after direct beam on the negative hemisphere of the chamber
|
|
- Move between the 2 limits and look for when the limit light flashes: They are about 10 degrees apart, a little after direct beam on the negative hemisphere of the chamber
|
|
|
- E1 = cal = limit reverse
|
|
- E1 = cal = limit reverse
|
|
|
- E2 = rm = limit forward
|
|
- E2 = rm = limit forward
|
|
|
- 
|
|
- 
|
|
|
- Move to a value more positive than E1
|
|
- Move to a value more positive than E1
|
|
|
- Set limit settings to normally closed for both directions
|
|
- Set limit settings to normally closed for both directions in: 
|
|
|
- once in between the 2 limits, click ncal 
|
|
- once in between the 2 limits, click ncal 
|
|
|
- software should reset tth to 0
|
|
- software should reset tth to 0
|
|
|
- reset soft limits
|
|
- reset soft limits
|
| ... | | ... | |