... | ... | @@ -11,13 +11,14 @@ To home tth from ![image](uploads/980f2f9c2666873734aa9fac9361a535/image.png): m |
|
|
- In pop up screen, enter IP address: 164.54.118.24
|
|
|
- Turn current on: ![image](uploads/32cfab7202eaab25ab23f014faf2b804/image.png) => Motor On + send
|
|
|
- Limit switches are close to the position where d4 is in direct beam. Check soft limits allow movement over that range: ![image](uploads/33c6c1d2a222252947b903e610567405/image.png)
|
|
|
- Set limit settings to ignore normally open x2: ![image](uploads/22033937627b72b32825c700066dc509/image.png)
|
|
|
- Set limit settings to ignore normally closed for both directions: ![image](uploads/22033937627b72b32825c700066dc509/image.png)
|
|
|
![image](uploads/650417152f86e41f8ac51341651efdca/image.png)
|
|
|
- move between the 2 limits: they about 10 degrees apart, a little after direct beam on the negative hemisphere of the chamber
|
|
|
- Move between the 2 limits and look for when the limit light flashes: They are about 10 degrees apart, a little after direct beam on the negative hemisphere of the chamber
|
|
|
- E1 = cal = limit reverse
|
|
|
- E2 = rm = limit forward
|
|
|
- ![image](uploads/1a6b0405009c84853cbd395acc9ad865/image.png)
|
|
|
- set limit settings to normally closed x2
|
|
|
- Move to a value more positive than E1
|
|
|
- Set limit settings to normally closed for both directions
|
|
|
- once in between the 2 limits, click ncal ![image](uploads/e40cef0f3899e5828930152ac30811f6/image.png)
|
|
|
- software should reset tth to 0
|
|
|
- reset soft limits
|
... | ... | |