How to home tth from ![image](uploads/980f2f9c2666873734aa9fac9361a535/image.png); note: make sure the .vi is running: ![image](uploads/f4bd29981acf9b15d811c2e3e3161eea/image.png)
|
|
|
## How to home tth from labview (preferred):
|
|
|
|
|
|
|
|
|
To home tth from ![image](uploads/980f2f9c2666873734aa9fac9361a535/image.png): make sure the .vi is running: ![image](uploads/f4bd29981acf9b15d811c2e3e3161eea/image.png)
|
|
|
|
|
|
- enter IP address: 164.54.118.24
|
|
|
- turn current on: ![image](uploads/32cfab7202eaab25ab23f014faf2b804/image.png) => Motor On + send
|
... | ... | @@ -31,14 +34,14 @@ ___ |
|
|
|
|
|
|
|
|
|
|
|
# Random assortment of notes about PI motors
|
|
|
## Random assortment of notes about PI motors
|
|
|
|
|
|
If KTH is shaking: make sure Integral (I) coeff at stanstill =< 0.002 or lower
|
|
|
|
|
|
PI stages are running left handed in the native software.
|
|
|
Dial positions in epics match PI software after the following homing procedure.
|
|
|
|
|
|
PI stages homing procedure:
|
|
|
PI stages homing procedure from the PI software (not recommended, see labview procedure above instead):
|
|
|
|
|
|
TTH (6449V31100 - 164.54.118.24): The reference switch is located around direct beam. Pick your direction depending on where TTH is: if TTH is on the outboard (inboard) hemisphere of the chamber, use positive (negative) limit.
|
|
|
Home without changing the max/min allowed position
|
... | ... | |