... | ... | @@ -99,6 +99,14 @@ To reboot controllers remotely (equivalent to power cycling the controller): |
|
|
WARNING: power cycle reset parameters! To make sure parameters are saved see below.
|
|
|
|
|
|
|
|
|
|
|
|
## Save parameters
|
|
|
|
|
|
- go to terminal
|
|
|
- save all
|
|
|
|
|
|
|
|
|
|
|
|
## Troubleshooting steps
|
|
|
|
|
|
- power cycle controllers
|
... | ... | @@ -125,12 +133,6 @@ however |
|
|
- Extra wires on kth/kap stages = thermocouple: max T =180deg ; keep T <160 deg
|
|
|
|
|
|
|
|
|
## Save parameters
|
|
|
|
|
|
- go to terminal
|
|
|
- save all
|
|
|
|
|
|
|
|
|
## Routine rotation service
|
|
|
|
|
|
- motor should be moved over their entire range on a regular basis (every week) to prevent sticky point
|
... | ... | |