| ... | ... | @@ -25,7 +25,19 @@ To home tth from : m |
|
|
|
- Set limit settings to ignore normally closed for both directions in: 
|
|
|
|
- Close LabView
|
|
|
|
|
|
|
|
## How to home KAP and KTH:
|
|
|
|
|
|
|
|
- Start PI software
|
|
|
|
- #Note that if an axis will not move, check soft limits
|
|
|
|
- Check your limits:
|
|
|
|
Set maximum allowed position to 100 (-100 in diffractometer coordinates, "User").
|
|
|
|
Set minimum allowed position to -210 (210 in diffractometer coordinates, "User").
|
|
|
|
Redefine TTH=0 in epics by scanning direct beam on d4 if beam is present.
|
|
|
|
- First zero KAP (417005588) stage. Starting from a position where the sample is facing up, the limit is located around -44
|
|
|
|
- To zero, open expanded view axis window and click on home with negative limit. When it reaches the limit, it will set the value to zero
|
|
|
|
- Before homing KTH (417005589), move KAP to 44 which will move the sample back to facing vertical.
|
|
|
|
- #Note: Watch the cooling braid during homing
|
|
|
|
- Open expanded axis view window and reference using negative limit. Is located around outboard 10CF window (ie KTH ~ -40). At limit it will set KTH motor to 0.
|
|
|
|
|
|
|
|
|
|
|
|
___
|
| ... | ... | @@ -159,7 +171,7 @@ please use the SMCHydraDemo Program. |
|
|
|
## Troubleshooting
|
|
|
|
|
|
|
|
- switch from close loop adaptive to open loop
|
|
|
|
- decrease velocity (for kth we used 1 instead of 1.5)
|
|
|
|
- decrease velocity (for kth we used 1 instead of 1.5)
|
|
|
|
- increase motor current; idealy, want to keep operating current under the following values:
|
|
|
|
- in vacuum: 0.7-0.8A
|
|
|
|
- in air: 1.2A<br>
|
| ... | ... | |