- increase motor current; ideally, want to keep operating current under the following values:
|
|
|
- in vacuum: 0.7-0.8A
|
|
|
- in air: 1.2A<br>
|
|
|
however
|
|
|
however we run a much higher current to begin with.
|
|
|
- swap motor cable at the controller
|
|
|
- measure resistance of the motor (unplug cable from Hydra): 4<span>Ω</span> between pin 1-2 & 3-4
|
|
|
|
... | ... | |