... | ... | @@ -83,7 +83,7 @@ ___ |
|
|
# Troubleshooting
|
|
|
|
|
|
|
|
|
## Remote reboot:
|
|
|
## Remote reboot
|
|
|
|
|
|
To reboot controllers remotely (equivalent to power cycling the controller):
|
|
|
|
... | ... | @@ -99,7 +99,7 @@ To reboot controllers remotely (equivalent to power cycling the controller): |
|
|
WARNING: power cycle reset parameters! To make sure parameters are saved see below.
|
|
|
|
|
|
|
|
|
## Troubleshooting steps:
|
|
|
## Troubleshooting steps
|
|
|
|
|
|
- power cycle controllers
|
|
|
- switch from close loop adaptive to open loop
|
... | ... | @@ -120,7 +120,7 @@ however |
|
|
- plug back motor, run in air, then pump and run in vacuum (remember to change current setting)
|
|
|
|
|
|
|
|
|
## Temperature:
|
|
|
## Temperature
|
|
|
|
|
|
- Extra wires on kth/kap stages = thermocouple: max T =180deg ; keep T <160 deg
|
|
|
|
... | ... | @@ -131,12 +131,12 @@ however |
|
|
- save all
|
|
|
|
|
|
|
|
|
## Routine rotation service:
|
|
|
## Routine rotation service
|
|
|
|
|
|
- motor should be moved over their entire range on a regular basis (every week) to prevent sticky point
|
|
|
|
|
|
|
|
|
### Pictures:
|
|
|
## Pictures
|
|
|
|
|
|
![PISoftware](uploads/08437da143c4ae5f2e61ab793334627f/PISoftware.jpg)
|
|
|
|
... | ... | |