## warning
|
|
|
- the vortex is sensitive to visible light so the lamp should be off and windows closed
|
|
|
- if you cycle power or turn on the light, you will need to restart the IOC (software).
|
|
|
- after turning on the vortex, wait 5 min for it to cool down
|
... | ... | @@ -10,7 +12,7 @@ To control the vortex, you only need the little Vortex tab located in the upper |
|
|
![image](uploads/b8cde90802840b8627e5dc0797fb65b0/image.png)
|
|
|
|
|
|
|
|
|
## To power on/off the vortex:
|
|
|
## to power on/off
|
|
|
|
|
|
On the screen press ![image](uploads/c452321c3f922b3043653038b09cba06/image.png) or ![image](uploads/21a7b5fefd526aa1c3325e5ffd0f5e58/image.png) <br>
|
|
|
|
... | ... | @@ -22,9 +24,9 @@ mcaoff() |
|
|
|
|
|
|
|
|
|
|
|
## To turn ON/OFF the IOC (software):
|
|
|
## to start software
|
|
|
|
|
|
**Note: Always turn power ON the vortex before starting the IOC.**<br>
|
|
|
**Note: Always turn power ON the vortex before starting the IOC (software).**<br>
|
|
|
The IOC can only be run from **Sneezy** (which is physically connected to the vortex via USB).<br>
|
|
|
To log into Sneezy, open a terminal:
|
|
|
|
... | ... | @@ -61,7 +63,7 @@ Vortex screen goes all white when the IOC is OFF. |
|
|
[sneezy] vortex_status
|
|
|
```
|
|
|
|
|
|
## Setting up the vortex parameters:
|
|
|
## setting up parameters
|
|
|
Once the IOC is ON, set the parameters as follow:
|
|
|
1. set the Preamp gain to 2.7 and then back to 2.5.
|
|
|
2. press ![image](uploads/63aee19d1a96b28be8659638e7e8a838/image.png)
|
... | ... | @@ -82,7 +84,7 @@ You can visualize the vortex spectrum in the Vortex ROIs tab on the bottom right |
|
|
To run in continuous mode, switch the drop down menu from ![image](uploads/82df6a63db833cf440b4a5d256800138/image.png) to a time that matches the vortex acquisition time (typically 1 or 2s).
|
|
|
|
|
|
|
|
|
## To take data:
|
|
|
## to take data
|
|
|
|
|
|
|
|
|
1. Add the vortex as detector so it is saved during the scan:
|
... | ... | |